Changes of hematological references depends on storage period and temperature conditions in rats and dogs

Abstract

Because changes in rat and dog hematological parameters according to storage conditions have been poorly documented, we sought to examine such changes. Blood analysis was performed using two hematology analyzers (ADVIA 2120i and Sysmex XN-V) after storage at room temperature and in cold storage for 5, 24, and 48 h, respectively. Interassay coefficients of variation for hematological parameters analyzed with the ADVIA 2120i and the XN-V showed similar. The levels of hematocrit, mean corpuscular volume, mean corpuscular hemoglobin, mean corpuscular hemoglobin concentration, and platelet (PLT) showed significant variations with time in blood samples of rats and dogs. The leukocyte subpopulation showed high variation with storage conditions. The data for leukocyte differential counts obtained using the ADVIA 2120i, XN-V, and a manual differential counting procedure showed good agreement for neutrophils and lymphocyte counts, but monocytes, eosinophils, and basophils showed differences between the procedures. In conclusions, most rat and dog hematological parameters showed minimal changes; however, some showed high variation with storage time and temperature, especially PLT and leukocyte subpopulations. In conclusion, when performing hematological analysis in dogs and rats, it will be exactitude to analyze blood samples in fresh condition and at least within 24 h in the cold storage.

References

  1. Bourgès-Abella NH, Geffré A, Deshuillers PL, Braun JP, Trumel C. Changes in hematology measurements in healthy and diseased dog blood stored at room temperature for 24 and 48 hours using the XT-2000iV analyzer. Vet Clin Pathol 2014; 43(1): 24–35.

    Article  Google Scholar 

  2. Gulati GL, Hyland LJ, Kocher W, Schwarting R. Changes in automated complete blood cell count and differential leukocyte count results induced by storage of blood at room temperature. Arch Pathol Lab Med 2002; 126(3): 336–342.

    PubMed  Google Scholar 

  3. Goyal VK, Kakade S, Pandey SK, Gothi AK, Nirogi R. Determining the effect of storage conditions on prothrombin time, activated partial thromboplastin time and fibrinogen concentration in rat plasma samples. Lab Anim 2015; 49(4): 311–318.

    Article  CAS  Google Scholar 

  4. Médaille C, Briend-Marchal A, Braun JP. Stability of selected hematology variables in canine blood kept at room temperature in EDTA for 24 and 48 hours. Vet Clin Pathol 2006; 35(1): 18–23.

    Article  Google Scholar 

  5. Matsuzawa T. Present status of animal clinical pathology examinations in the Japanese Pharmaceutical Manufacturers Association. Toxicol Pathol 1992; 20(3 Pt 2): 528–533.

    Article  CAS  Google Scholar 

  6. Wood BL, Andrews J, Miller S, Sabath DE. Refrigerated storage improves the stability of the complete blood cell count and automated differential. Am J Clin Pathol 1999; 112(5): 687–695.

    Article  CAS  Google Scholar 

  7. Zandecki M, Genevieve F, Gerard J, Godon A. Spurious counts and spurious results on haematology analysers: a review. Part II: white blood cells, red blood cells, haemoglobin, red cell indices and reticulocytes. Int J Lab Hematol 2007; 29(1): 21–41.

    CAS  PubMed  Google Scholar 

  8. Bauer N, Nakagawa J, Dunker C, Failing K, Moritz A. Evaluation of the automated hematology analyzer Sysmex XT-2000iV ™ compared to the ADVIA ® 2120 for its use in dogs, cats, and horses. Part II: Accuracy of leukocyte differential and reticulocyte count, impact of anticoagulant and sample aging. J Vet Diagn Invest 2012; 24(1): 74–89.

    Article  Google Scholar 

  9. England JM, Rowan RM, van Assendelft OW, Bull BS, Coulter W, Fujimoto K, Groner W, Richardson-Jones A, Klee G, Koepke JA, Lewis SM, McLaren CE, Shinton NK, Tatsumi N, Verwilghen RL. Recommendations of the International Council for Standardization in Haematology for Ethylenediaminetetraacetic Acid Anticoagulation of Blood for Blood Cell Counting and Sizing. International Council for Standardization in Haematology: Expert Panel on Cytometry. Am J Clin Pathol 1993; 100(4): 371–372.

    Article  Google Scholar 

  10. Schoorl M, Schoorl M, Oomes J, van Pelt J. New fluorescent method (PLT-F) on Sysmex XN2000 hematology analyzer achieved higher accuracy in low platelet counting. Am J Clin Pathol 2013; 140(4): 495–499.

    Article  Google Scholar 

  11. Wada A, Takagi Y, Kono M, Morikawa T. Accuracy of a New Platelet Count System (PLT-F) Depends on the Staining Property of Its Reagents. PLoS One 2015; 10(10): e0141311.

    Article  Google Scholar 

  12. Meintker L, Ringwald J, Rauh M, Krause SW. Comparison of automated differential blood cell counts from Abbott Sapphire, Siemens Advia 120, Beckman Coulter DxH 800, and Sysmex XE-2100 in normal and pathologic samples. Am J Clin Pathol 2013; 139(5): 641–650.

    Article  Google Scholar 

  13. Tan BT, Nava AJ, George TI. Evaluation of the Beckman Coulter UniCel DxH 800, Beckman Coulter LH 780, and Abbott Diagnostics Cell-Dyn Sapphire hematology analyzers on adult specimens in a tertiary care hospital. Am J Clin Pathol 2011; 135(6): 939–951.

    Article  Google Scholar 

  14. Zandecki M, Genevieve F, Gerard J, Godon A. Spurious counts and spurious results on haematology analysers: a review. Part I: platelets. Int J Lab Hematol 2007; 29(1): 4–20.

    CAS  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Jin Seok Kang.

Rights and permissions

This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (https://doi.org/creativecommons.org/licenses/by-nc/3.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Lee, JM., Kang, J.S. Changes of hematological references depends on storage period and temperature conditions in rats and dogs. Lab Anim Res 32, 241–248 (2016). https://doi.org/10.5625/lar.2016.32.4.241

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.5625/lar.2016.32.4.241

Keywords